CPR Medical Devices, Inc. CPR Medical Devices, Inc.
CPR Medical Devices, Inc.
CPR Medical Devices Inc. address
E-mail CPR Medical Devices Inc.
 
Menu Tools: Open All  Close All  Locate  Move Previous  Move Next 

Home
Oxylator® Product Series
Overview
Oxylator® EM-100
Introduction
Brochure (PDF)
Operating Manual (PDF)
Algorithms and Charts
Components



Disassembly for Cleaning



Comparison of Features among Ventilation Devices


How to Use Effectively
Photographs
Publications


Patents, Approvals, and Clearances
Case Studies



St. Gallen Cantonal Hospital, Switzerland (PDF)



Hospital Princeps d'Espanya Bellvitge, Barcelona, Spain



Royal Victoria Hospital and McGill University, Montreal, QC, Canada



Montérégie's EMS System, Longueil, QC, Canada



NTV a Nederlands Tijdschrift Voor Anesthesi- medewerkers, Netherlands



Helicopter Emergency Medical Services, University Hospital Rotterdam, Netherlands



University of Massachussetts Medical Center, Worcester, MA, U.S.A.



Emergency Scientific Medical Center, Yerevan, Armenia


CPR Medical Devices Inc., Toronto, ON, Canada
Testimonials


Carter County Emergency & Rescue Squad, Inc., Elizabethton, TN, U.S.A.


University of Massachusetts Medical Center, Worcester, MA, U.S.A.


U.S. Department of Veteran Affairs, Dublin, GA, U.S.A.


Croft Rescue Squad, Spartanburg, SC, U.S.A.


Lenoir Memorial Hospital, Kinston, NC, U.S.A.


Dunn Rescue Squad, Inc., Dunn, NC, U.S.A.

Jefferson County EMS, Dandridge, TN, U.S.A.
Oxylator® FR-300
Introduction
Brochure (PDF)
Operating Manual (PDF)
Photographs
Publications


Patents, Approvals, and Clearances
Case Studies


St. Elisabeth Hospital, Tilburg, NL

University of Massachussetts Medical School, Worcester, MA, U.S.A.
Oxylator® EMX
Introduction
Brochure (PDF)
Operating Manual (PDF)
Photographs
Publications

Patents, Approvals, and Clearances
Distributors
Demonstration Videos
Case Studies
Oxylator® EM-100


St. Gallen Cantonal Hospital, Switzerland (PDF)


Hospital Princeps d'Espanya Bellvitge, Barcelona, Spain


Royal Victoria Hospital and McGill University, Montreal, QC, Canada


Montérégie's EMS System, Longueil, QC, Canada


NTV a Nederlands Tijdschrift Voor Anesthesi- medewerkers, Netherlands


Helicopter Emergency Medical Services, University Hospital Rotterdam, Netherlands


University of Massachussetts Medical Center, Worcester, MA, U.S.A.


Emergency Scientific Medical Center, Yerevan, Armenia

CPR Medical Devices, Inc., Toronto, ON, Canada
Oxylator® FR-300

St. Elisabeth Hospital, Tilburg, NL
University of Massachussetts Medical School, Worcester, MA, U.S.A.
Testimonials
Oxylator® EM-100

Carter County Emergency & Rescue Squad, Inc., Elizabethton, TN, U.S.A.

University of Massachusetts Medical Center, Worcester, MA, U.S.A.

U.S. Department of Veteran Affairs, Dublin, GA, U.S.A.

Croft Rescue Squad, Spartanburg, SC, U.S.A.

Lenoir Memorial Hospital, Kinston, NC, U.S.A.

Dunn Rescue Squad, Inc., Dunn, NC, U.S.A.
Jefferson County EMS, Dandridge, TN, U.S.A.
Distributors
Brochures
Oxylator® EM-100 (PDF)
Oxylator® FR-300 (PDF)
Oxylator® EMX (PDF)
Operating Manuals
Oxylator® EM-100 (PDF)
Oxylator® FR-300 (PDF)
Oxylator® EMX (PDF)
Publications
Oxylator® EM-100
Oxylator® FR-300
Oxylator® EMX
Demonstration Videos
News and Newsworthy

Helicopter Emergency Medical Services, University Hospital Rotterdam, Netherlands
REA 2000, Ostschweizer Bildungsaustellung, St. Gallen, Switzerland
Downloads
Brochures
Operating Manuals
Publications
Demonstration Videos
FAQs
Acknowledgments
Site Map
Contact

Features
Oxylator® outperforms bag-valve, in the European Journal of Anaesthesiology Oxylators® reviewed in the Journal of Emergency Medical Services
Oxylator® bests bag-valve in peer-review studies Oxylators® reviewed in JEMS magazine
Internal study, 1995

Studies of the Oxylator® EM-100 Resuscitation System


CPR Medical Devices, Inc.
161 Don Park Road, Markham, Ontario, Canada


Introduction | Protocol | Methods | Results | Discussion | Conclusions


Introduction

The EM-100 Oxylator® is a compact, hand-held resuscitator/inhalator intended for use by emergency response personnel whenever a patient's ventilatory ability has been compromised. The EM-100 may be used as a resuscitator or, for a patient who is breathing spontaneously but is in need of supplemental oxygen enrichment, as an inhalator. It is a stand-alone unit, requires a source of appropriately-regulated oxygen to function (pressure of 50 psi), and offers the care-giver great flexibility in responding to the patient's needs.

The Oxylator® contains a valve that delivers compressed 100% oxygen at a maximum flow rate of 40 litres per minute from a pressure source of 50 psi. The system also permits the adjustment of it's maximum inspiratory pressures from a value of 25 cmH2O to 50 cmH2O.

The system has four different operation modes:

1. Manual mode
2. Manual mode with the addition of a baseline pressure (PEEP)
3. Automatic cycling mode with a baseline pressure (PEEP)
4. Inhalation mode which provides oxygen enriched air

The third mode, automatic cycling, was used in our investigations. This mode was used in order to study the efficacy of the EM-100 in delivering adequate ventilation simultaneously with continuous chest compressions.

The EM-100 Oxylator® system meets all of the standard recommendations and guidelines for oxygen powered resuscitation devices published in JAMA 268, 2199-2241, 1992.

^ Menu


Protocol

Investigation 1: using the Oxylator® EM-100 resuscitation device (automatic cycling mode) and a simulated human mannequin.

A modified 'foam-filled' human simulated mannequin (Adam) was used in these experiments. A cavity was formed inside of the mannequin by removing some of the underlying foam material. The cavity was later filled with 2 (2 litres) 'Penlon' anaesthesia test lungs. Their expansion limit was controlled by covering the open end cavity with a thick card board.

The mannequin's endotraceal tube was connected to the respirometer first and then to the Oxylator® EM-100. The mannequin is equipped with a one way valve that is capable of creating an airway blockage, if the head of the mannequin is not appropriately tilted. The Oxylator® was set at the automatic cycling mode, at a pressure of 50 cmH2O. The EM-100 gives both a visual an audible indication of such an obstruction by 'clicking' rapidly.

Investigation 2: using the Oxylator® EM-100 resuscitation device (automatic cycling mode), a simulated human mannequin and a Thumper™ (cardiopulmonary resuscitator system) set at 5:1 ratio (5 compressions to 1 ventilation cycle).

Same as for Investigation 1, but with the addition of the Thumper™. The Thumper™ is a cardiopulmonary resuscitator system that is placed on the patient's sternum and compressions occur at a rate of 80 compressions per minute. Between each 5 continuous chest compressions, the Thumper™ stops and delivers 1 complete cycle of ventilation. The Thumper™ is also capable of delivering continuous chest compressions with no interruptions.

The Thumper™ was also set to deliver 80 chest compressions per minute, at a vertical displacement of the chest, of about 1.5-2 inches. (The contact surface area of the Thumper™ pad, was 3 cm x 2 cm.)

Investigation 3: using the Oxylator® EM-100 resuscitation device (automatic cycling mode), a simulated human mannequin and a Thumper™ (cardiopulmonary resuscitator system) in continuous chest compressions mode.

Same conditions as in Investigation 2, except in this case the Thumper™ was allowed to deliver continuous chest compressions at a rate of 80 compressions per minute in conjunction with the automatic cycling of the Oxylator®.

The three investigations were designed to determine the efficacy of the Oxylator® EM-100 in delivering adequate oxygenation in cardiopulmonary resuscitation attempts, in three different settings:

1. By itself
2. With the Thumper™ set at 5:1 ratio (five chest compressions to one breath)
3. With the Thumper™ set at continuous chest compression mode

In evaluating the Oxylator's functional characteristics, the following